Certainty
Chapters
27, 3, 32
Clinical implication
Comparator
Design
Consensus statement
Identifier
Intervention
Limitations
Main result
Population
Verified
Year
2026
International multidisciplinary group of ECLS experts; recommendations and standardised criteria for in vitro device testing according to international norms, so that results from different research groups become comparable.
What Chapter 27 takes from it:
- ISO 7199 specifies an inlet condition of SO2 65 ± 5%, at haemoglobin 120 g/L — the third and lowest of the three standardised inlet saturations now recorded in this book for rated-flow testing (against 70% in the ELSO Red Book and ISCCM Manual and 75% in the practical-guide textbook and, internally, in Bartlett 2020).
- The minimum physiological requirement: an oxygenator should exchange at least 50 mL of oxygen per litre of blood, derived as 1.34 mL/g × 120 g/L × (95% − 65%) plus the dissolved term.
- The corresponding carbon dioxide minimum of 40 mL per litre for an adult on a mixed diet, from a respiratory quotient of 0.8; 50 mL/L in neonates and children on a predominantly carbohydrate diet with a quotient near 1.
- Gas exchange should be reported as millilitres of gas per litre of blood across a range of blood flow rates, which is independent of device type or size and allows oxygenators to be compared directly.
- Gas exchange testing must use whole blood, because the oxygen bound to haemoglobin far exceeds the dissolved fraction; stored blood must be evaluated for red cell age effects on binding and release.
Why it matters to this book: it is the document that makes the rated-flow discrepancy consequential — two devices tested at different inlet saturations are not comparable, and a bedside clinician comparing an observed outlet saturation to a printed rated flow is making exactly that comparison.
Read as indexed full-text passages. DOI not independently resolved.